Subcutaneous versus transvenous implantable defibrillator: An updated meta-analysis.

Published
November 20, 2020
Journal
Heart rhythm
PICOID
cea91a63
DOI
Citations
40
Keywords
Appropriate therapy, Complications, Implantable cardioverter-defibrillator, Inappropriate therapy, Subcutaneous, Transvenous
Copyright
Copyright © 2020 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.
Patients/Population/Participants

patients with an indication for ICD without the need for pacing

Intervention

S-ICD

Comparison

TV-ICD

Outcome

composite of clinically relevant device-related complications and inappropriate shock

Abstract

P
I
C
O

Implantable cardioverter-defibrillator (ICD) placement is a well-established therapy for prevention of sudden cardiac death. The subcutaneous implantable cardioverter-defibrillator (S-ICD) was specifically designed to overcome some of the complications related to the transvenous implantable cardioverter-defibrillator (TV-ICD), such as lead complications and systemic infections. Evidence on the comparison of S-ICD vs TV-ICD are limited. The purpose of this study was to conduct an updated meta-analysis comparing S-ICD vs TV-ICD. Electronic databases were searched for studies directly comparing clinical outcomes and complications between S-ICD and TV-ICD. The primary outcome was the composite of clinically relevant complications (lead, pocket, major procedural complications; device-related infections) and inappropriate shocks. Secondary outcomes included death and the individual components of the primary outcome. Thirteen studies comprising 9073 patients were included in the analysis. Mean left ventricular ejection fraction was 40% ± 10%; 30% of patients were female; and 73% had an ICD implanted for primary prevention. There was no statistically significant difference in the risk of the primary outcome between S-ICD and TV-ICD (odds ratio [OR] 0.80; 95% confidence interval [CI] 0.53-1.19). Patients with S-ICD had lower risk of lead complications (OR 0.14; 95% CI 0.06-0.29; P <.00001) and major procedural complications (OR 0.18; 95% CI 0.06-0.57; P = .003) but higher risk of pocket complications (OR 2.18; 95% CI 1.30-3.66; P = .003) compared to those with TV-ICD. No significant differences were found for the other outcomes. In patients with an indication for ICD without the need for pacing, TV-ICD and S-ICD are overall comparable in terms of the composite of clinically relevant device-related complications and inappropriate shock.

Similar article map

CEO: Hwi-yeol YunCOO: Jung-woo ChaeCTO: Sangkeun Jung
Location: 204, W6, Chungnam National University, 99, Daehak-ro, Yuseong-gu, Daejeon, Republic of Korea
Tel: 042-821-7328E-mail: webmaster@lilac-co.kr
Copyright © 2024 by LiLac. All Rights Reserved.